wtw VWR MU 6100 H Operating Manual page 46

Table of Contents

Advertisement

Ion concentration
Displaying the cali-
bration data memory
Calibration evalua-
tion
Calibration record
(example)
46
The calibration records of the last calibrations are available in the menu,
/ Calibration / Calibration data storage.
Menu item
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to USB
After calibrating, the meter automatically evaluates the calibration.
Display
Error
Error elimination (see section 15 W
..., page 89)
DO IF
MU 6100 H
Ser. no. 12345678
CALIBRATIONISE
18.01.2013 08:09:10
Standard 1
Standard 2
Voltage 1
Voltage 2
Temperature 1
Temperature 2
Ion type
Slope
Sensor
_____________________________________
Setting/func-
Explanation
tion
-
Displays the calibration record.
Further options:
 Scroll through the calibration
 Output the displayed calibra-
 Quit the display with
 Switch directly to the mea-
-
Outputs the calibration records to
the interface.
Calibration
record
+++
Error
HAT TO
0.010 mg/l
0.020 mg/l
38.5 mV
58.0 mV
24.0 øC
24.0 øC
Ag
64.7 mV
+++
MU 6100 H
<
><
records with
tion record to the interface with
<F2>
/[USB output].
<F1>
<OK>
[Back] or
.
sured value display with
<MODE>
.
Magnitude of the slope [mV]
50.0 ... 70.0 or 25.0 ... 35.0
< 50 or > 70
or
< 25 or > 35
ba77077e06
<OK>
>
.
/
05/2021