Resetting The System Settings - wtw VWR MU 6100 H Operating Manual

Table of Contents

Advertisement

MU 6100 H
ba77077e06
05/2021
settings with the Reset function:
Setting
Kal.-Intervall
Measured parameter
Cell constant (C)
(calibrated)
Cell constant (C)
(set)
Temperature compensation
Reference temperature
Temperature coefficient (TC) of the
linear temperature compensation
TDS multiplier
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
<OK>
<F1>
press the
(or

11.3.2 Resetting the system settings

The following system settings can be reset to the default status:
Setting
Language
Beep
Baud rate
Output format
Decimal separator
Contrast
Illumination
Switchoff time
Stability control
Temperature unit
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu in the measured value display, press the
_
<OK
>
<F1__>
(or
Default settings
150 d
ϰ
0.475 cm
(nominal cell constant of the conduc-
tivity measuring cell last calibrated)
0.840 1/cm
nLF
25 °C
2.000 %/K
1.00
/[Menu]) key.
Default settings
English
On
4800 Baud
ASCII
Dot (xx.x)
50 %
Auto
1 h
On
°C
/[Menu]) key.
Settings
-1
-1
or 0.840 cm
73

Advertisement

Table of Contents
loading

This manual is also suitable for:

665-0311665-0312

Table of Contents