MU 6100 H
ba77077e06
05/2021
pH
9.958
AnyCal
15.02.2014
26.
Set the nominal buffer value for the measured temperature with
<
><
>
.
27.
Accept the set calibration value with
The calibration display for the next buffer appears (voltage display).
<MODE>
28.
Use
to finish calibration or
Continue calibrating using the next buffer with
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
6.2.5
Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
Determined
values
1-point
Asy
2-point
Asy
Slp.
3-point to
Asy
5-point
Slp.
You can display the slope in the units, mV/pH or % (see section
11.1.1 S
ETTINGS FOR P
3
Buffer
24.8
°C
14:15
<OK>
Displayed calibration data
Zero point = Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
Zero point = Asy
Slope = Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated by lin-
ear regression.
H
MEASUREMENTS
pH value
.
<OK>
.
, page 60).
35