4.8.4 Settings for IDS conductivity sensors

The following settings are possible for IDS conductivity sensors:
Calibration record (display)
Calibration interval
Reference temperature
Temperature compensation
TDS factor
Unit of the temperature
Automatic Stability control
The settings are made in the Measurement menu for the measured
parameter, conductivity. To open the settings, display the required
measured parameter in the measured value display and press the
<MENU/ENTER> key. After completing the settings, switch to the mea-
sured value display with <M>.
The possible settings are individually displayed for each sensor. Below
the setting menu is displayed for two IDS sensors (TetraCon 925,
LR325/01).
Default settings are printed in bold.
Menu item
Calibration /Cali-
bration record
Calibration /Cali-
bration interval
Measurement /
Measuring cell /
Type
Possible
Description
setting
-
Displays the calibration record of
the last calibration.
1 ... 150 ...
Calibration interval for the IDS con-
999 d
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring window.
Measuring cell used
cal
Measuring cells whose cell con-
stant is determined by calibration
in the KCL control standard solu-
tion.
Calibration range:
0.450 to 0.500 cm
The currently valid cell constant is
displayed in the status line.
man
Freely adjustable cell constant in
the range 0.450 ... 0.500 cm
Operation
-1
-1
.
65