wtw VWR MU 6100 H Operating Manual page 67

Table of Contents

Advertisement

MU 6100 H
ba77077e06
05/2021
Menu item
Measurement /
Measuring cell /
Type
Measurement /
Measuring cell /
Cell const. man
Measurement /
Measuring cell /
Cell const. LR325/
01
Measurement /
Temp. comp. (TC)
/ Method
Measurement /
Temp. comp. (TC)
/ Linear coeff.
Possible
Explanation
setting
Measuring cell used
Measuring cells whose cell constant
Cal
is determined by calibration in the
KCL control standard solution.
Calibration ranges:
0.450 ... 0.500 cm
0.800 ... 0.880 cm
The currently valid cell constant is
displayed in the status line.
C 0.100
Measuring cell C 0.100,
nominal cell constant
0.100 cm
The cell constant can be adjusted in
the range 0.090 ... 0.110 cm
Measuring cell C 0.010,
C 0.010
nominal cell constant
0.010 cm
The cell constant is permanently set.
Any measuring cells with freely
man
adjustable cell constants.
Display and setting option of the cell
0,250 ...
constant of any measuring cells
(man).
Display and setting options for the cell
0.090 ...
constant of the measuring cell
0.100 ...
-1
C 0.100.
0.110 cm
Procedure for temperature compen-
nLF
sation (see section 11.1.9 T
Lin
Off
TURE COMPENSATION
This setting is only available for the
measured parameters,
Coefficient of the linear temperature
0.000 ...
compensation.
2.000 ...
10.000 %/K
This setting is only available when the
linear temperature compensation is
set.
Settings
-1
and
-1
-1
.
-1
.
-1
.
EMPERA
, page 68).
ϰ
ρ
and
-
.
67