wtw VWR MU 6100 H Operating Manual page 76

Table of Contents

Advertisement

Data memory
Settings
Starting the auto-
matic storing func-
tion
Energy saving mode
([Eco mode])
76
You can configure the automatic data storing function with the following
settings:
Menu item
ID number
Interval
Duration
To start the automatic storing function, select Continue with
<OK>
confirm with
. The meter switches to the measured value display.
pH
0
6.949
0d03h00min
15.02.2014
Eco mode
14:15
1 Remaining storing duration
2 Graphical display of the storing duration
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura-
tion.
If the automatic storage function is activated, only the following
keys are active: Softkeys,
The other keys and the automatic shut-off function are deactivated.
If the automatic storing function is active, the meter provides an energy saving
mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
Possible set-
ting
1 ... 10000
1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min
1 min ... x min
14
25.0
°C
<MODE>
Explanation
ID number for the dataset series.
Storing interval.
The lower limit of the storing inter-
val can be restricted by the num-
ber of free memory locations.
The upper limit is restricted by the
storing duration.
Storing duration.
Specifies after which time the
automatic data storing should be
terminated.
The lower limit of the storing dura-
tion is restricted by the storing
interval.
The upper limit is restricted by the
number of free memory locations.
<
><
1
2
_
<STR
>
,
and
ba77077e06
MU 6100 H
>
and
<On/Off>
.
05/2021

Advertisement

Table of Contents
loading

This manual is also suitable for:

665-0311665-0312

Table of Contents