Calibration Interval - wtw VWR MU 6100 H Operating Manual

Table of Contents

Advertisement

Settings
Setting the calibra-
tion interval
62
No.
Buffer set *
5
TEC 2
Technical buffers
6
Merck 1*
7
Merck 2 *
8
Merck 3 *
9
Merck 4 *
10
Merck 5 *
* Brand names or trade names are trademarks of their respective owners protected
by law.
The buffers are selected in the menu, pH /
Buffer (see 11.1.1 S

11.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
1.
Open the menu for measurement settings with
2.
In the Calibration / Calibration interval menu, set the calibration interval
<
><
>
with
.
pH values
4.010
7.000
10.011
4.000
7.000
9.000
1.000
6.000
8.000
13.000
4.660
6.880
9.220
2.000
4.000
7.000
10.000
4.010
7.000
10.000
H
ETTINGS FOR P
MEASUREMENTS
MU 6100 H
at
25 °C
20°C
20°C
20°C
20°C
25 °C
<OK>
/ Calibration /
,
60).
PAGE
<OK>
.
ba77077e06
05/2021