wtw VWR MU 6100 H Operating Manual page 33

Table of Contents

Advertisement

MU 6100 H
Continuing with two-
point calibration
ba77077e06
05/2021
7.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
8.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
pH
AnyCal
AR
15.02.2014
9.
Wait for the end of the measurement with stability control or accept the
calibration value with
The calibration display for the setting of the nominal buffer value appears.
pH
7.000
AnyCal
15.02.2014
10.
Set the nominal buffer value for the measured temperature with
<
><
>
.
11.
Accept the set calibration value with
The calibration display for the next buffer appears (voltage display).
12.
If necessary, finish the calibration procedure as a single-point calibration
<MODE>
with
.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.
13.
Thoroughly rinse the electrode with deionized water.
<OK>
.
7.000
Buffer
0.0
mV
24.8
°C
14:15
<OK>
.
1
Buffer
24.8
°C
14:15
<OK>
pH value
.
33