MU 6100 H
Measuring ranges,
resolution
(depending on the
sensor)
Accuracy (± 1 digit)
Correction functions
ba77077e06
05/2021

3.1.2 Oxi

Variable
Concentration [mg/l]
Saturation [%]
D.O. partial pressure
[mbar]
T [°C]
Variable
Concentration [mg/l]
Saturation [%]
D.O. partial pressure
[mbar]
T [°C] / temperature sensor
NTC 30
PT 1000
Temperature
compensation
Salinity correction
Air pressure correc-
tion
The accuracy values specified here apply exclusively to the meter.
The accuracy of the D.O. sensors has also to be taken into
account.
Measuring range
0 ... 20.00 (0 ... 20.0)
20.0 ... 90.0
(20 ... 90)
0 ... 200.0 (0 ... 600)
0 ... 600
0 ... 200.0 (0 ... 1250)
0 ... 1250
0 ... 50,0
Accuracy
± 0.5 % of measured value
at ambient temperature of +5 °C ... +30 °C
± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
± 0.5 % of measured value
at ambient temperature of +5 °C ... +30 °C
± 0.1
± 0.1
Accuracy better than 2 % at 0 ... +40 °C
0 ... 70.0 SAL
Automatic through integrated
pressure sensor in the range of 500 ... 1100
mbar
Technical data
Resolution
0.01 (0.1)
0.1 (1)
0.1 (1)
1
0.1 (1)
1
0.1
11