15.3 Conductivity

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
The measured value is outside the measuring range.
Cause
– The measured value is outside
the measuring range of the meter
Cause
– Measuring cell contaminated
– Calibration solution not suitable
Remedy
– Use a suitable measuring cell
Remedy
– Clean cell and replace it if nec-
essary
– Check the calibration solutions
ba77077e06