Thoroughly rinse the electrode with distilled water.
8.
Immerse the electrode in standard solution 1.
9.
When calibrating without temperature sensor:
 Measure the temperature of the standard solution using a thermom-
eter.
<F2>
 Use
/[
<
> <
 Use
<F2>
 Use
/[
10.
Set the concentration of the standard solution with
<OK>
press
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
ISE
AR
11.
Wait for the end of the AutoRead measurement or accept the calibration
<OK>
value with
The calibration display for the next standard solution appears.
1
Standard
0.0
mV
24.8
°C
0.010
mg/l
15.02.2014
12:12
] to select the setting of the temperature.
>
to set the temperature.
] to select the setting of the concentration.
0.01 mg/l
0.0
mV
24.8
°C
15.02.2014
12:12
.
Ion concentration
<
> <
>
and
43