Table of Contents

Advertisement

MU 6100 H
ba77077e06
05/2021
O
2
0.88
25.1
OxiCal
15.02.2014
4.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. A progress bar is displayed and
the display of the measured parameter flashes.
5.
Wait for the end of the measurement with stability control or accept the
calibration value with
The calibration record is displayed and output to the interface.
<F1>
6.
Using
/[Continue] or
The meter displays a D.O. saturation of approx. 101.7 % in the air cali-
bration vessel.
9.2.5
Calibrating with a comparison measurement (OxiComp)
For this calibration procedure, the Comparison meas. setting must be set to On
in the Calibration menu.
Before calibrating via a comparison measurement, the sensor
should be calibrated in the air calibration vessel.
1.
Connect the D.O. sensor to the meter.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with
O
2
7.9
25.1

OxiComp

15.02.2014
°C
08:00
<OK>
.
<OK>
.
<OK>
, switch to the measured value display.
<CAL>
.
mg/l
°C
08:00
Dissolved oxygen
51

Advertisement

Table of Contents
loading

This manual is also suitable for:

665-0311665-0312

Table of Contents