MU 6100 H
Accuracy (± 1 digit)
Dimensions
Weight
Mechanical structure
Electrical safety
Test certificates
Ambient
conditions
ba77077e06
05/2021
Variable
ϰ and ρ / temperature compensation
None (Off)
Nonlinear (nLF)
Linear (lin)
SAL
/ range
0.0 ... 42.0
TDS [mg/l]
T [°C] / temperature sensor
NTC 30
PT 1000
The accuracy values specified here apply exclusively to the meter.
The accuracy of the measuring cell has also to be taken into
account.
Further data are given in the documentation of your sensor.
3.2
General data
ca. 180 x 80 x 55 mm
Approx. 0.4 kg
Type of protection
Protective class
CE
Storage
Operation
Admissible relative
humidity
Accuracy
Temperature of the test
sample
± 0.5 %
± 0.5 %
0 °C ... +35 °C
according to EN 27 888
± 0.5 %
+35 °C ... +50 °C
enhanced nLF function
± 0.5 %
+10 °C ... +75 °C
± 0.1
+5 °C ... +25 °C
± 0.2
+25 °C ... +30 °C
± 0.5 %
± 0.1
± 0.1
IP 67
III
-25 °C ... +65 °C
-10 °C ... +55 °C
Yearly mean: < 75 %
30 days/year: 95 %
Other days: 85 %
Technical data
13