wtw VWR MU 6100 H Operating Manual page 37

Table of Contents

Advertisement

MU 6100 H
Calibration record
(example)
ba77077e06
05/2021
record.
Display
Clean the electrode according to the elec-
trode operating manual
Error
Error elimination (see section 15 W
..., page 89)
DO IF
MU 6100 H
Ser. no. 11292113
CALIBRATIONpH
15.02.2014 15:55
AutoCal TEC
Buffer 1
Buffer 2
Buffer 3
Voltage 1
Voltage 2
Voltage 3
Temperature 1
Temperature 2
Temperature 3
Slope
Asymmetry
Sensor
etc...
Calibration
record
+++
-15 ... +15
++
-20 ... <-15
or
>+15 ... +20
+
-25 ... <-20
or
>+20 ... +25
-
-30 ... <-25
or
>+25 ... +30
Error
<-30 oder
>+30
HAT TO
4.01
7.00
10.01
184.0 mV
3.0 mV
-177.0 mV
24.0 °C
24.0 °C
24.0 °C
-60.2 mV/pH
4.0 mV
+++
Zero point
Slope [mV/
[mV]
pH]
-60.5 ... -58.0
>-58.0 ... -57.0
-61.0 ... <-60.5
or
>-57.0 ... -56.0
-62.0 ... <-61.0
or
>-56.0 ... -50.0
<-62.0 oder
>-50.0
pH value
37