wtw VWR MU 6100 H Operating Manual page 34

Table of Contents

Advertisement

pH value
Continuing with
three- to five-point
calibration
34
14.
Immerse the electrode in the second buffer solution.
15.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
16.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
17.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
The calibration display for the setting of the nominal buffer value appears.
pH
4.035
AnyCal
15.02.2014
18.
Set the nominal buffer value for the measured temperature with
<
><
>
.
19.
Accept the set calibration value with
The calibration display for the next buffer appears (voltage display).
20.
Finish the calibration procedure as a two-point calibration with
The calibration record is displayed.
21.
Thoroughly rinse the electrode with deionized water.
22.
Immerse the electrode in the next buffer solution.
23.
When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
24.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
The calibration display for the setting of the nominal buffer value appears.
<OK>
.
2
Buffer
24.8
°C
14:15
<OK>
<OK>
.
MU 6100 H
.
<MODE>
ba77077e06
<OK>
.
.
<OK>
.
05/2021