Ion concentration
Continuing with two-
point calibration
44
ISE
Ag
12.
Thoroughly rinse the electrode with distilled water.
13.
Immerse the electrode in standard solution 2.
14.
When calibrating without temperature sensor:
 Measure the temperature of the standard solution using a thermom-
eter.
<F2>
 Use
/[
<
> <
 Use
<F2>
 Use
/[
15.
Set the concentration of the standard solution with
<OK>
press
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
ISE
AR
16.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with
The calibration display for the next standard solution appears.
ISE
Ag
2
Standard
0.0
mV
24.8
°C
0.020
mg/l
15.02.2014
12:12
] to select the setting of the temperature.
>
to set the temperature.
] to select the setting of the concentration.
0.02 mg/l
8.9
mV
24.8
°C
15.02.2014
12:12
<OK>
.
3
Standard
20.5
mV
24.8
°C
0.050
mg/l
15.02.2014
12:12
MU 6100 H
<
> <
>
and
ba77077e06