6.6
Storage
If the system is not used for a longer time period (more than a few days), perform the
following actions:
• Remove the flow kit (see
• Clean the system (see
• If the system is to be moved, the system power and air supply must be switched off.
Note:
ÄKTA readyflux Operating Instructions 29175550 AG
Section 5.8 Procedures after a run, on page
General cleaning on page
Although the system can be stored at temperatures in the range
-25°C to 40°C, room temperature is preferable.
Flow kits should not be stored at temperatures below 0°C.
NOTICE
Do not leave a flow kit mounted on the system for an excessively
long time. The tubing walls may stick together where the tubing is
squeezed, possibly causing high pressure alarms when running the
system.
NOTICE
Do not use 1 M NaOH (or higher) for more than 2 hours. Flush after-
wards with a suitable rinsing liquid to remove the NaOH.
204).
211).
6 Maintenance
6.6 Storage
215