5.5.1

About the system installation test

Introduction
The installation test checks that the system is correctly installed and that the modules
are functioning properly. An installation test takes approximately 60 minutes.
Note:
Note:
Prerequisites
• Make sure that the flow kit type mounted on the system matches the flow kit type
selected in the software.
• Make sure that the flow kit optional components and service are selected in compo-
nent view.
• Make sure that the latest installation test method is used.
• Make sure that the flow kit is installed according to the instructions in the
ÄKTA readyflux flow kit installation wizard.
• If using a ¼ inch flow kit, the pump gain factor has been set to the default value, see
Section 4.7 Install the pump upgrade kit, on page
• Make sure that the Bagkart trolley or the weighing scale is calibrated and tared.
• Make sure that the feed inlets have been primed to remove air from the feed pump.
• The pH electrode needs to be wetted for approximately 2 hours before use, to
achieve accurate readings.
Description of the system installation
test
The following table describes the different stages of the system installation test.
Stage
1
2
3
4
ÄKTA readyflux Operating Instructions 29175550 AG
An installation test should always be performed after installing a new
system or new parts.
System installation test protocol covers all the components that are part of
the flow path, i.e., the complete testing of a flow kit plus configuration.
Description
Individual valves are opened and closed.
Their pressure regulator is checked for warnings during individual valves
open/close test.
The conductivity, temperature and UV sensors are tested by pumping a
solution of 1.06% (v/v) acetone and 1.0 M sodium chloride through the
flow kit. If the UV and conductivity readings are within the specified
range, the sensors pass the test.
The pH sensor is tested for properties of demineralized water.
5.5 System installation test
5.5.1 About the system installation test
112.
5 Operation
189