Note:
ÄKTA readyflux Operating Instructions 29175550 AG
Action
Turn the black levers on the pinch valves 180 degrees to close the safety
locks on all pinch valves:
• 3 valves on the transfer side of the system
• 3 valves on the front of the system
• 3 valves on the permeate side of the system
Turn the FLOW KIT INSTALL - RUN - INTEGRITY TEST selector switch to
the RUN position.
FLOW KIT
RUN
INSTALL
CAUTION
Pinch hazard from unused valves. When using the
NFF flow kit, not all valves on the system are in use. When
switching from INSTALL to RUN mode or in case of
power failure, valves can change from an open to a
closed state. Never put fingers or any objects into the
unused pinch valve openings.
Note:
Software can only start when RUN is selected.
Click the END button in the UNICORN System Control dialogue to
complete the flow kit installation wizard.
INTEGRITY
TEST
5 Operation
5.2 Flow kit installation
5.2.4 Install the NFF flow kit
166