6.1

Safety precautions

Maintenance
ÄKTA readyflux Operating Instructions 29175550 AG
WARNING
Disconnect power. Always disconnect power from the instru-
ment before performing any maintenance task.
WARNING
LOCK OUT / TAG OUT (LOTO)! Before any maintenance or
decommissioning work is performed on the system, make sure
that:
• the system is empty and depressurized.
• the system is disconnected from process feed, electrical power
and pneumatic supply.
• the system is prevented from accidentally becoming re-ener-
gized during maintenance.
• the system is clearly tagged as taken out of operation.
• all process wetted areas are clean and decontaminated.
WARNING
Only personnel authorized by Cytiva may perform service, installa-
tion, and maintenance of components inside the electrical cabinet
of the system.
WARNING
Use only approved parts. Only spare parts and accessories that
are approved or supplied by Cytiva may be used for maintaining or
servicing the product.
CAUTION
Connectors on the cabinet, such as connectors for pH, conduc-
tivity, network, etc., that are not used, should be plugged to prevent
cleaning liquid from entering the connector.
6 Maintenance

6.1 Safety precautions

208