5.5.2
Run a system installation test
Equipment and chemicals
The following equipment and chemicals are required for the system installation test:
• CFF plus flow kit containing conductivity, pH and UV sensors on permeate section
• T-connector of the same size as the flow kit, O-rings, TC clamps, and flexible tubing
• 30 L of purified water
• For ⅜ inch and ½ inch flow kits, a 20 L bag
For ¼ inch flow kits, a 5 L bag
• Container to hold approximately 20 L of waste solution
• Test solution: 10 L of 1.06% (v/v) acetone and 1.0 M sodium chloride in a suitable
container. See also the Test solution preparation protocol template on the system
CD.
Note:
Note:
Prepare for the system installation
test
Follow the steps below to prepare for the system installation test.
Step
1
2
3
4
5
Note:
ÄKTA readyflux Operating Instructions 29175550 AG
The test solution must be freshly prepared before an installation test is
performed, to prevent the acetone from evaporating.
Make sure that the purified water and the test solution are equilibrated to
room temperature.
Action
Connect the feed outlet to the flow kit permeate and retentate inlets with a
T-connector.
Connect ⅜" i.d. flexible tubing on XV-082, XV-083, XV-071, XV-031 and
XV-034 (P-203 or AE-131) outlet and place in a container with purified
water.
For a description on how to make the connection, see
connections of the flow kit using TC end connectors on page
Open the lid of the Permeate pump (P-203).
Place containers next to the system and use only required tubing lengths to
connect flow kit and containers.
Follow the instructions on the screen.
The following warnings can appear during the test and can be ignored:
• The "UV led low light intensity" warning appears if UV is enabled and the
lamp is not switched on.
5 Operation
5.5 System installation test
5.5.2 Run a system installation test
Connect the end
147.
191