Calibration
Introduction
This section describes how to calibrate the components.
Calibration of 2D bag
To perform a calibration of the 2D bag, a pre-calibrated hanging weight of known
weight must be used.
Follow these steps to perform a calibration:
Step
1
2
3
Calibration of flow sensor FE-141
The flow sensor must be filled with water without air bubbles and must be still before
performing the zero calibration.
Follow these steps to perform a calibration:
Step
1
2
3
Calibration of pumps P-201, P-202
and P-203
Calibration can only be performed in Manual Instructions and must be calibrated
according to the instructions below.
Note:
ÄKTA readyflux Operating Instructions 29175550 AG
Action
Start a run.
Go to System Control →Systems →Calibrate →Cal_WE176.
Follow the instructions displayed.
Action
Start a run.
Go to System Control →Manual →Special →FE141_AutoZero.
Follow the instructions displayed.
The pump upgrade kit must be installed to use P-201 with ¼ inch flow kits.
The pump gain factors must be set in the software before the first time that
a ¼ inch flow kit is used. See
on page 112
for instructions.
Section 4.7 Install the pump upgrade kit,