Step
9
ÄKTA readyflux Operating Instructions 29175550 AG
Action
Attach a drive adapter to a torque wrench. Tighten the screws diagonally on
the pump pressure plate to 8 Nm ( for ½ inch or ⅜ inch flow kits) or 3 Nm (for
¼ inch flow kits). Verify correct torque circularly.
5 Operation
5.2 Flow kit installation
5.2.3 Install the CFF flow kit