Voltage Selection and Fuse Rating

5.3
The Magstim Rapid
~ 50Hz and 240V ± 10% ~ 60Hz. The voltage selector on the rear panel of the Main Frame and if
relevant, the Magstim Plus
indicated in the table below. The PSU, however, is not voltage selectable and is manufactured as
either a 230V (240V) ± 10% unit, or a 115V ± 10% unit. The two are not interchangeable.
Supply Voltage
115V ±10%~60Hz
230V ±10%~50Hz
240V ±10%~60Hz
T denotes timed or antisurge fuses. Fast acting fuses are not recommended.
H denotes high breaking capacity (ceramic fuse).
Note: Only replace fuses with those of the correct rating, and only operate the instrument with the
voltage selector set at the appropriate voltage. Failure to do so may cause serious damage to the
device(s).
Fuse Replacement:
5.3.1
1. The Magstim Rapid² must be disconnected from the mains by disconnecting the power cord from
the PSU.
2. Insert the tip of a small blade screwdriver or similar tool into the slot (2) (refer to diagram 5.2) and
gently lever up the retaining lip of the fuse holder. The fuse holder will then slide out.
3. Replace the damaged fuse(s) with one of the correct rating (See above to ensure the correct fuse
is replaced) and locate and push home the tray ensuring the retaining lip clicks into place.
Changing the Voltage Selector Setting:
5.3.2
1. The Magstim Rapid² must be disconnected from the mains by disconnecting the power cord from
the PSU.
2. Remove the fuse tray as instructed above,
3. Using a pair of narrow nosed pliers, grip the voltage selector unit on one of the metal connection
plates and remove it from the Power Entry Module.
© The Magstim Company Limited
2
may be operated from supplies in the range of 115V ± 10% ~ 60Hz, 230V ± 10%
1
along with the associated fuses must correspond to the supply voltage as
Voltage
Main Frame
Selector
Fuse Rating
Setting
115V
T1.25AL
230V
T1.25AL
230V
T1.25AL
Figure 5.2: Power Entry Module
46
Magstim
1
Quantity
Plus
Fuse
Rating
2
T250mAL
2
T250mAL
2
T250mAL
(1)
(2)
Magstim Rapid²
Operating Manual
Quantity
2
2
2
MOP03-EN-03