Equipment manufactured by The Magstim Company Limited is fully guaranteed, covering both materials and workmanship, for a period of one year from the date of shipment. The Magstim Company Limited reserves the right to perform guarantee services in its factory, at an authorised repair station, or at the customer’s installation.
Magstim recommends that a standard questionnaire is completed to screen a patient’s suitability prior to using any Magstim product on that patient and concurs with the recommendations given in the paper titled “Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research”...
Damage: If there are any signs of damage to the Coils, or if any parts are damp or wet, the coil must not be used*. If damaged, the coil should be returned to the Magstim Company Limited for servicing and repair (see Section 7.2 for contact details).
Page 7
The use of ear protection is recommended for the operator and adult patients. Ear protection is mandatory for children. Seizures: cortical magnetic stimulation runs the risk of inducing seizures. Magstim concurs with the advice given in the paper titled “Safety, ethical considerations and application guidelines for the use of TMS in clinical practice and research”...
Coil connector Coil head Coil handle Figure 3.1: Example Coil illustration (3190-00 – D70mm Remote Control Coil) Double 70mm Remote Control Coil This coil has a figure-of-eight winding in the coil head. Current flow is in opposing directions in each half of the coil, which is indicated by the current direction label (see Figure 3.2).
Page 10
3.4.1.4 Down Button* (4) When used with the Magstim 200² or Bistim² engaging the down button will reduce the selected parameter by one unit. If the button is held, the parameter will continue to decrease until the button is released or the minimum value is reached.
The Magstim Company Limited for servicing and repair. (See section 5.4 for servicing details). If cleaning is required, please follow the instructions provided in Section 5.3.
Operation 4.3.1 Use of Remote Control Coils with Magstim 200² When the coil is connected to the stimulator and ready to use, remove the coil from the coil holder of the right hand side of the stimulator*. The coil will show the output power of the system on the multi-function display*.
Page 14
4.3.2 Use of Remote Control Coil with Magstim Bistim² When the coil is connected to the stimulator and ready to use, remove the coil from the coil holder of the right hand side of the master unit*. The coil multi-function display* will show one of the values listed below.
Page 15
Note: Ensure that the system is disarmed prior to disconnecting the stimulating coil from the stimulator 4.3.4 Use of the Standard Coil with Magstim 200², BiStim² and Rapid² When the coil is connected to the stimulator and ready to use, select the desired stimulator output power/ protocol using the stimulator UI.
Continued use in this condition will eventually cause the contacts to deteriorate and loose electrical connection, which will cause damage to the system. Repair is a specialised procedure and can only be undertaken by The Magstim Company Limited or an authorised service centre. If any pins show damage, however slight, contact The Magstim Company Limited service department for further advice or to arrange a return (see Section 7.2 for...
Generation Coils reaches the end of its serviceable life, it should not be disposed of in general waste. The Magstim Company Limited should be contacted (see Section 7.2 for contact details) for advice on its disposal in compliance with the appropriate environmental regulations.
100% 100% Power Power Power When used with a Magstim 200², at an ambient temperature of 20°C. Table 6.1: Coil specifications 6.2.1 Electric Field Plot Figure 6.1: Approximate electric field profile for 90mm Remote Coil and 90mm Standard Coil measured using a Magstim Rapid²...
Page 21
Magstim 2nd Generation Coils Operating Manual Figure 6.2: Approximate electric field profile for D70mm Remote Coil measured using a Magstim Rapid² 6.2.2 Magnetic Field Plot Figure 6.3: Approximate Magnetic Field Profile for 90mm Remote Coil and 90mm Standard Coil measured using a Magstim Rapid²...
Magstim 2nd Generation Coils Operating Manual Figure 6.4: Approximate Magnetic Field Profile for D70mm Remote Coil measured using a Magstim Rapid² Environmental Conditions Operating temperature: 5°C to 40°C Transport and storage temperature: -19°C to 60°C Operating, transport and storage relative humidity:...
APPENDIX B - TMS MEDICAL SCREENING QUESTIONNAIRE The Magstim Company Limited takes your health and safety very seriously and as such, it is necessary to point out that there is a remote possibility of adverse effects occurring as a result of receiving Transcranial Magnetic Stimulation (TMS). Such effects can include a risk of seizure, headache, pain and nausea in particular susceptible individuals.
Page 26
Magstim 2nd Generation Coils Operating Manual The Magstim Company accepts no responsibility for adverse reactions resulting from (TMS) where false or inaccurate medical information is supplied. Magstim have strict confidentiality policies and we confirm that your information will be kept securely and will not be passed on to any other third party.
The Magstim Rapid System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Magstim Rapid System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF...
Page 31
If the measured filed strength in the location in which the Magstim Rapid System is used exceeds the applicable RF compliance level above, the Magstim Rapid System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Magstim Rapid System.
Page 32
The Magstim 200 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Magstim 200 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF...
Page 35
If the measured filed strength in the location in which the Magstim 200 is used exceeds the applicable RF compliance level above, the Magstim 200 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such...