MAGSTIM Rapid2 Operating Manual page 40

Hide thumbs Also See for Rapid2:
Table of Contents

Advertisement

Magstim Rapid²
Operating Manual
In order to save the traces, it is necessary to go back to Single Pulse Mode and save the data as
before. The information will remain on each trace window until a new trace is generated. Then it will
be overwritten by a new one.
Setup: Selection of this option will return the user to the Repetitive Setup screen.
Single Pulse: Selecting this option will return the user to the Single Pulse Mode screen.
Options Menu: Selection of this option will return the user to the Options Menu screen.
System Options:
4.4.5
User Configuration: Selection of this option
will call up the User Configuration Screen.
System Details: Selection of this option will call
up the System Details Screen.
System Configuration: Selection of this option
will call up the System Configuration Screen.
Options Menu: Selection of this option will
return the user to the Options Menu Screen.
Figure 4.24: System Options Menu Window
4.4.5.1
User Configuration
This screen allows the user to set system
parameters, which are stored in an area of the
UI's memory which is battery powered.
Volume:
This allows the user to set the
screen selection beep volume by selecting the
window and using the Rotary Control Knob.
Contrast : This allows the user to adjust the
TFT contrast option from 1 - 10, in steps of 1.
Coil Switch Option: This gives the user the
choice to use the coil safety switch or not each
time the system is armed. When ON the
System will prompt the user on a Run Screen
Figure 4.25: User Configuration Window
to state if they wish to use the coil safety switch.
Note: If the coil switch option is set to ON, each time the system is armed the user will have a
YES/NO option on the screen to ignore the switches before the system can fire. When the 'YES'
button is selected a small yellow button will appear in the top right hand corner of the system status
panel to indicate 'ignore' option is active.
It is important that those in the vicinity of the system are made aware that the system is in this mode of
operation, to ensure that inadvertent triggering does not occur.
© The Magstim Company Limited
37
MOP03-EN-03