Magstim Rapid²
Operating Manual
Channels 1, 2 and Both: Selecting Channel 1,
Channel 2 and Both determines whether both
traces are shown simultaneously, or individually.
Coil Status Panel: The Coil Status panel will
show the temperature and type of coil.
System Status Panel: The System Status
panel shows the system status. If a charge
delay has been set, a stopwatch symbol (
)
will be displayed.
Trace Window: The Trace Window will be
updated when stimulation occurs. The maximum
update rate for the trace window is, 4 times a
Figure 4.21: Repetitive Mode (Burst) Window
second, depending on the amount of information
to be displayed.
If the amplitude cursors are active then the last position set for each cursor in Single Pulse Mode will
be displayed. The cursor position will remain fixed until Single Pulse Mode is re-entered.
In order to save the trace, it is necessary to go back to Single Pulse Mode and save the data as
before. The information will remain on each trace window until a new trace is generated, when the
new trace will overwrite the previous trace.
Resume: This Option is available if the system is armed and the previous session was not completed.
Selecting Resume will return the session to the point at which it was stopped at the beginning, at the
start of the next cycle regardless of when the last cycle was stopped.
Setup: Selection of this option will return the user to the Repetitive Setup screen.
Single Pulse: Selecting this option will return the user to the Single Pulse Mode screen.
Options Menu: Selection of this option will return the user to the Options Menu screen.
Session Run Mode:
4.4.4
4.4.4.1
Session Run Mode: Setup
This screen allows the user to set up a series of
individual lines, each with different Power,
Frequency, Duration and Number of Pulses
parameters. See section 4.4.2 for limits. If
Session Setup is accessed from Single Pulse
Mode, the protocol intensity value established in
Single Pulse Mode will be carried over into the
power window.
To alter the settings, touch the window of the
required setting. A dark blue margin will appear
within the selected window. The settings can be
quickly altered using the Rotary Control Knob.
To alter the settings of a subsequent line, touch
the window of the required setting on that
Figure 4.22: Session Mode Setup Window
particular line. Subsequent lines cannot be
selected if the required settings have not been set on the previous line.
© The Magstim Company Limited
35
MOP03-EN-03