5.3
Voltage Selection and Fuse Rating
The Magstim Bistim
2
system is preset at the factory to operate at 110V-120V or 220V-240V A.C., as
appropriate. The following are the fuse ratings for both the master and slave units:
Supply Voltage
110V-120V-
220V-240V-
Qty
2
2
Fuse Size
20 x 5mm
20 x 5mm
Rating
T10A 250V
T5A250V
T denotes timed or antisurge fuses. Fast acting fuses are not recommended.
The fuses may be changed by inserting the tip of a small blade screwdriver or similar tool into the slot
below the arrow to lever the retaining lip enabling the fuse tray to be pulled out. Replace the blown
fuse(s) with one of the correct rating and locate and push home the tray ensuring the retaining lip
clicks into place.
The voltage may be selected by removing the fuse tray, rotating the selector and replacing the cover.
A legend will indicate the voltage selected when the fuse cover is replaced.
Only replace fuses with those of the correct rating, and only operate the instrument with the
voltage selector set at the appropriate voltage. Failure to do so may cause serious damage to
the device(s).
5.4
Cleaning and Disinfecting
The Magstim Bistim
2
system and accessories cannot be sterilised; therefore, do not allow them to
become contaminated with bodily fluids. They may be cleaned using an isopropyl-moistened cloth but
ensure that they have dried thoroughly before use.
5.5
Servicing
The Magstim Bistim
2
contains no user serviceable parts. Servicing of Magstim Bistim
2
system and
accessories must only be carried out by The Magstim Company Limited or one of its authorised
service centres. To arrange a return or for further information, contact the service department at The
Magstim Company Limited (see Section 7.2 for contact details).
5.6
Device Lifetime
5.7
'--~---·
bzr
~
The lifetime of the Magstim Bistim
2
system is defined as being five years from the date of shipment.
The Magstim Company Limited will support the products for the duration of their lifetime.
Disposal
When a Magstim Bistim
2
system or accessory reaches the end of its serviceable life, The Magstim
Company Limited should be contacted (see Section 7.2 for contact details) to arrange for its disposal
in compliance with the appropriate environmental regulations.