SECTION 4
Note: to avoid problems with interference, the stimulating coils should not be used in the vicinity of
any equipment not in compliance with the EMC Standard EN 60601-1-2, including mobile phones.
Note: Please ensure you are familiar with all sections of this operating manual prior to following these
operating instructions. Refer to section 3 for identification and descriptions of Rapid² connecting parts
and functions.
Note: Please consult any labelling and information accompanying stimulating coils and other
accessories for safety and use information regarding these devices.

Preparation

4.1
1.
At the start of each session the operator must check the Magstim Rapid² and any accessories
used for any signs of external damage and to identify any cleaning required*. If any cracks are
visible in the housing, or there is damage to any of the cables, the items must not be used and
should be returned to The Magstim Company Limited for servicing and repair. If cleaning is
required, please follow the instructions in Section 5.4 of this Operating Manual.
Note: The contacts in the coil connector, and the stimulator coil output socket should be checked
regularly for any signs of tarnishing or burning. See Section 5.2 for an example of contact
burning.
Rapid² Set-Up
4.2
Note: The Magstim Rapid² system should be assembled following the instructions provided within
this section to ensure continued compliance of IEC 60601-1:2005.
Note: ensure not to position the Magstim Rapid² system, so it is difficult to disconnect the power
lead.
Before using the instrument, ensure that the correct fuses are fitted and that the voltage selector
is set to suit the country's supply voltage (this statement applies to the Rapid
Magstim Plus
1.
Place the system in a room near multiple power sockets and ensure easy access to and around
the equipment. Make sure to stack the system as shown in Section 3.
2.
Label the entrances to the room to warn of the presence of strong magnetic fields and to exclude
those wearing pacemakers and/or electronic implants.
3.
Attach the coil holder to the mainframe by removing the two blanking plugs from the threaded
inserts on the right-hand side of the mainframe.
Insert the two M6x12 hexagon bolt into the mounting holes on the coil holder. Place the coil
holder against the right-hand side of the mainframe so that the bolts locate in the threaded
inserts. (See Figure 4.1).
Using the supplied hexagon key tighten the bolts so as to secure the coil holder to the right-hand
side of the main frame. Do not over tighten the bolts as this may damage the cases of the units.
© The Magstim Company Limited

OPERATING INSTRUCTIONS

1
only). See Section 5.3 for further information.
19
Magstim Rapid²
Operating Manual
2
Main Frame and
MOP03-EN-03