MAGSTIM Rapid2 Operating Manual page 30

Hide thumbs Also See for Rapid2:
Table of Contents

Advertisement

Note: The Charge Delay Function is still available when the MEP Pod is not connected.
To alter the settings, touch the window next to the required setting. A dark blue margin will appear
within the selected window. The settings can be quickly altered using the Rotary Control Knob.
Windows with a white background are not selectable.
4.4.1.1
The Power Panel is selected by pressing the
Power button.
button turns green.
Stimulator Output: To adjust the system power
level, touch the Stimulator Output window on
the Power Panel. The power level is set to the
default value of 30%.
increased, or decreased, in 1% increments by
using the Rotary Control Knob. Each complete
360º rotation changes the output level by
approximately 20%.
%MT: The %MT (Motor Threshold) window,
displays the percentage of stimulator output
which is to be transferred to the Protocol Intensity window. The %MT cannot be adjusted to give more
than the 100% in the Protocol Intensity window.
Enhanced: This option enables the power level to be increased up to 110% power. This option is
only available in Single Pulse Mode.
Note: If Repetitive Mode or Session Mode is accessed via Single Pulse Mode, the Protocol intensity
value will be carried over into these modes.
Note: Enhanced mode is not available when using an Air Film Coil (P/N 3950-00)
4.4.1.2
The Charge Delay Panel is selected by pressing
the Charge Delay button. When Charge Delay
is selected, the button turns green.
Status: Selects the status window to turn the
charge delay ON or OFF. When the status
window displays ON, a stopwatch symbol is
displayed in the System Status panel. The
default setting is OFF.
Duration: When the charge delay window
displays OFF, the duration window is not
selectable and can not be adjusted. When the
charge delay displays ON, the duration window
allows the duration of the charge delay to be
adjusted between 1ms and 10000ms in 1ms
increments using the Rotary Control Knob.
© The Magstim Company Limited
Display,
Save Traces
Power Panel
When Power is selected, the
The power can be
Charge Delay Panel
Figure 4.12: Power Panel Window
Figure 4.13: Charge Delay Panel Window
27
Magstim Rapid²
Operating Manual
MOP03-EN-03