MAGSTIM Rapid2 Operating Manual page 35

Hide thumbs Also See for Rapid2:
Table of Contents

Advertisement

Magstim Rapid²
Operating Manual
Duration (Charge Delay): When the Charge Delay is OFF, the duration window associated with the
charge delay is not selectable and can not be adjusted. A value of 0ms is displayed in the Duration
window. With the Charge Delay is ON, this window will display any previously set charge delay since
power-up. Selecting this window allows the duration of the charge delay to be adjusted using the
Rotary Control Knob between 1ms and 10000ms in 1ms increments. The default duration on power-up
is 0ms.
Charge Delay Operation: When incrementing the charge delay duration, the 'Wait Time' will
automatically be adjusted if required. The charge delay duration is included as part of the overall wait
time (this is shown graphically in brackets underneath the wait time).
The charge delay becomes active after the last pulse of a pulse train. After the last pulse, the system
will not begin re-charging until the specified charge delay duration has elapsed. Any attempt to trigger
the system during this charge delay period will be ignored.
Information Window: Appears at the bottom of the screen. The Information Window provides
information such as 'Coil OK to run,' and alerts the user to actions required, protocols that are unable
to be completed, or equipment which needs to be attached.
Print: For Print operations see Section 4.4.8
Save Session: Save Session takes the user to the File Maintenance screen (See section 4.4.6)
Options Menu: Selection of this option will return the user to the Options Menu screen.
Run Session: This screen takes the user to the Repetitive Mode screen in order to run a session.
4.4.2.2
Repetitive Run Mode: Standard
Session Details Panel: The Session Details panel shows a pictorial representation of the train
currently running, and shows the settings established in Repetitive Setup and any associated charge
delay duration if appropriate.
Channels 1, 2 and Both: Selecting
Channel
1,
Channel
2
or
both
determines whether both traces are
shown simultaneously or individually.
Coil Status Panel: The Coil Status
panel will show the temperature and type
of coil.
System Status: The System Status
panel shows the system status. If a
charge delay has been set, a stopwatch
symbol (
) will be displayed.
Figure 4.19: Repetitive Mode (Standard) Window
Trace Window: The Trace window will
be updated when a stimulation occurs. The maximum update rate for the trace window is, 4 times a
second, depending on the amount of information to be displayed.
If the amplitude cursors are active then the last position set for each cursor in Single Pulse Mode will
be displayed. The cursor position will remain fixed until Single Pulse Mode is re-entered
In order to save the traces it is necessary to return to Single Pulse Mode and save the data as before.
The information will remain on each trace window until a new trace is generated, when the new trace
will overwrite the previous trace.
© The Magstim Company Limited
32
MOP03-EN-03