115V Rapid² Systems:
Standby:
During Discharge
Input Fuse(s) Rating:
Rapid² mainframe:
Plus¹ unit:
Equipment should be used with the line cord provided. Where a fused line cord is used, the
appropriate fuse should be fitted.
General Specifications:
Rapid² Main Frame (3012-00):
Dimensions:
Weight:
Stacking limit:
Rapid² Single PSU (3013-00):
Dimensions:
Weight:
Stacking limit:
Rapid² Dual PSU (3014-00):
Dimensions:
Weight:
Stacking limit:
Rapid² Plus¹ (4040-00):
Dimensions:
Weight:
Stacking limit:
Rapid² UI (3022-00):
Dimensions:
Weight:
Software:
Base Software:
UI Controller Software:
MEP Pod Software:
Please note: All specifications are subject to alteration.
Accessory Specifications:
Stimulating Coils:
Compatibility:
MEP Pod:
Class Type:
Channels:
CMMR:
Dynamic Range:
© The Magstim Company Limited
All Systems:
20VA
Standard Rapid²:
2.3kVA
Super Rapid²:
4.6kVA
Rapid² Plus¹:
6.9kVA
2 x T1.25AL, 250V 20x5mm for 115V-240V~
2 x T250mAL, 250V 20x5mm for 115V-240V ~
460mm x 160mm x 375mm
13.1kg
4 Units stacked on top of each other
460mm x 160mm x 375mm
13.2kg
4 Units stacked on top of each other
460mm x 160mm x 375mm
23.4kg
4 Units stacked on top of each other
460mm x 160mm x 375mm
15.7kg
4 Units stacked on top of each other
360mmx x 240mm x 280mm
2.6kg
Version 9.0
(Software ref: 3056)
Version 11.0
(Software ref: 3473)
Version 2.0
(Software ref: 3519)
See Appendix E
Class 1, BF
2
> 90dB
± 30mV ±10%
51
Magstim Rapid²
Operating Manual
MOP03-EN-03

Advertisement

Table of Contents
loading

Table of Contents