Magstim Rapid²
Operating Manual
Protocol description.
This is the string that appears in the patient protocol description window on the UI and must not
exceed 46 characters. The string can contain any character except a comma.
Comments.
This is the string that appears in the patient comments window on the UI and must not exceed 273
characters. The string can contain any character except a comma.
Train.
This line contains the parameters displayed on the repetitive set up screen on the UI, (Standard
mode). The individual parameter value ranges are as follows:
Power 0 - 100% at a resolution of 1%.
Frequency 1 - 100 Hz at a resolution of 0.1Hz up to 30.0Hz, and at 1.0Hz thereafter.
Note the settable power to frequency values is determined by the system configuration, see user
manual for Rapid system for more details.
Duration 0.1 - 600.0 seconds at a resolution of 0.1 seconds, up to 30.0s, and at 1.0s thereafter.
Number of pulses 1 – 60000 at a resolution of 1.
Wait time 1.0 – 540.0 seconds at a resolution of 0.1 seconds.
Number of trains 1 – 999 at a resolution of 1.
Note all time periods are specified in multiples of tenths of a second i.e. if a time period of 6.5
seconds is required then a value of 65 is entered between the commas.
Burst.
This line contains the parameters displayed on the repetitive set up screen on the UI, (Burst mode).
The individual parameter value ranges are as follows:
Power: 0 - 100% at a resolution of 1%.
Frequency : 1 - 100 Hz at a resolution of 1Hz.
Note the settable power to frequency values is determined by the system configuration, see user
manual for Rapid system for more details.
Number of pulses: 1 – 10 at a resolution of 1.
Burst frequency: 1 – 10 Hz at a resolution of 1Hz.
Number of bursts: 1 – 500 at a resolution of 1.
Cycle time: 1.0 – 500.0 seconds at a resolution of 0.1 seconds.
Number of cycles: 1 – 999 at a resolution of 1.
Note all time periods are specified in multiples of tenths of a second i.e. if a time period of 6.5
seconds is required then a value of 65 is entered between the commas.