Cleaning And Disinfecting - MAGSTIM Rapid2 Operating Manual

Hide thumbs Also See for Rapid2:
Table of Contents

Advertisement

4. Orient the voltage selector unit, such that the desired voltage is facing out from the Power Entry
Module.
5. Re-insert the voltage selector unit into the Power Entry Module.
6. Replace the fuse holder, ensuring that the retaining lip is properly home, and verify that the voltage
window (1) displays the correct voltage value.
Backup Batteries
5.3.3
Battery status can be checked on the user configuration screen and when low, is reported on the main
Options Menu screen.
The batteries in the UI will require to be replaced every 2 years. To do so:
1. Orient the voltage selector unit, such that the desired voltage is facing out from the Power Entry
Module.
2. Unplug the UI connector and turn the UI upside down,
3. Remove the 4 cross-head screws visible, and detach the plastic bottom of the UI.
Note: The UI has a small capacitor which will maintain the internal settings for about 20 seconds once
the old batteries are removed. It is important that the batteries are replaced within the 20 seconds if
data is not to be lost.
4. Before removing the old batteries, ensure you have the correct replacements, and that you are clear
about the orientation of the batteries.
5. Once inserted, replace the bottom of the UI and fix using the screws. Do not over-tighten.
Note: Do not use re-chargeable batteries. When spent, the batteries should be disposed of under the
appropriate environmental regulations
Note: It is advisable that the batteries be removed from the UI if the equipment is likely to be un-used
for a long period of time.

Cleaning and Disinfecting*

5.4
The Magstim Rapid² exterior, UI, PSU, stimulating coil, MEP pod and MEP pod patient cables,
footswitch and
contaminated with bodily fluids. They may be cleaned using a cloth moistened with isopropyl alcohol,
however ensure that the Magstim Rapid² and accessories have dried thoroughly before use.
It is the responsibility of the user to ensure the Magstim Rapid² is cleaned when necessary.
Note: Do not clean or wipe the touch screen with anything abrasive as it will cause permanent
damage
Note: The Magstim Company Ltd recommends that a lint free cloth moistened with 70% isopropyl
alcohol should be used.
© The Magstim Company Limited
accessories cannot be sterilised; therefore, do not allow them to become
47
Magstim Rapid²
Operating Manual
MOP03-EN-03