File Maintenance Window
4.4.6
The file maintenance window allows the user to
enter a file name and store protocols. These may
be generic protocols which are chosen for a patient.
The protocols may be stored internally, or may be
stored on a memory card. (See section 4.4.6.2)
Keyboard: The keyboard is touch sensitive. A file
name needs to be selected in order to edit it. Select
the Local Store screen and use the Rotary control
knob to scroll between files.
View: Selection of 'View' reloads the data stored in
the selected file located in the local store window.
The file name will be displayed in the top right hand
Figure 4.28: File Maintenance Window
corner of the screen.
New: Selection of 'New' allows the user to create a new file in the local store window. The file name
entered must be between 1 and 18 characters long, and different from any other file name.
Open: Selection of 'Open' will reload the data stored in the selected file. Once the file is opened the 'Open'
button will change to 'Close' to enable the file to be closed. The file name will be displayed in the top right
hand corner of the screen.
Rename: Selection of this option will allow the user to enter a new name for a chosen File Name. Use the
keyboard to make changes to the file name and press save (on the keyboard) to confirm.
Delete: Selection of this option will allow the user to delete the selected file from the local store window
after a 'Yes' confirmation.
Export: Export allows the user to export the selected File name in the Local Store window to the multi-
media card. (See section 4.4.6.2)
Import: Import transfers the selected file from the multimedia card, to the Local Store window.
Print: For Print operations See Section 4.4.8.
Patient Details: Selection of this option will call up the Patient Details Screen. See Section 4.4.6.1
Back: Selection of this option returns the user to the previous screen.
Options Menu: Selection of this option will return the user to the Options Menu Screen.