MAGSTIM Rapid2 Operating Manual page 34

Hide thumbs Also See for Rapid2:
Table of Contents

Advertisement

Repetitive Mode: Standard
4.4.2
4.4.2.1
The Repetitive Setup Screen allows the user to
set the following adjustable parameters to the
desired values to enable a session to be run. If
Repetitive Setup is accessed from Single Pulse
Mode, the protocol intensity value established in
Single Pulse Mode will be carried over into the
power window.
To enter Standard Mode, select the 'Standard'
button on the Repetitive Setup screen which
allows the listed adjustable parameters to be
set.
To alter the settings, touch the window next, or
below, the required setting. A dark blue margin
will appear within the selected window. The
settings can be quickly altered using the Rotary Control Knob. Windows with a white background can
not be altered.
The status of the charge delay and duration are synchronised between Single Pulse Mode and
Repetitive Mode. That is, any changes made to the charge delay settings in Repetitive Mode will be
reflected in Single Pulse Mode.
Power: Select the power window and use the Rotary Control Knob to adjust the System Power Level.
The power can be increased, or decreased, in 1% increments.
Frequency: Selecting the Frequency window allows the user to set the frequency of the pulses to a
value between 0 and F
frequency can be up to 100Hz depending on the power level set and the system configuration. A table
detailing the available power/ frequency combinations is included in Appendix
Duration (Train): Selecting the Duration window allows the user to set the duration of the pulse train
to between 0.1 up to a maximum of 600 seconds, in 0.1 second increments up to 30 seconds, and in
1.0 second increments thereafter.
frequency settings and the connected coil (see Appendix E for coil compatibility information). On
entering the duration and frequency, the number of pulses is automatically calculated by the UI and
displayed.
Number of pulses: Selecting the Number of Pulses window allows the user to set the number of
pulses in a train to between 1 and 60000 in steps of 1. The number of pulses cannot be set to exceed
the maximum allowed duration. If the frequency is altered, the system will automatically adjust the
duration to fit the current number of pulses.
Wait Time: Selecting the 'Wait Time' allows the user to set the wait time to between 1 and 540
seconds, in 0.1 second steps. However, if the minimum wait time required for the selected protocol is
greater than that selected, the UI will display a message 'Wait period too short, auto-set to minimum'.
Select 'YES' to set the wait delay to the minimum allowed for the chosen protocol. Select 'NO' to reset
the protocol parameters manually.
No. of Trains: Selection of this option allows the user to set the number of times that the specified
train runs to a value between 1 and 999 in steps of 1. The number of trains cannot be set to exceed
the maximum total number of pulses of 65000.
Charge Delay: Select the 'Charge Delay' window to turn the Charge Delay on and off. The default
setting is OFF.
© The Magstim Company Limited
Repetitive Setup: Standard
Hz in 0.1Hz increments up to 30Hz, 1.0Hz thereafter. The maximum
Max
Duration (Train)
Mode Selection
Buttons
Information
Window
Figure 4.18: Repetitive (Standard) Setup Window
The maximum duration allowed depends on the power and
31
Magstim Rapid²
Operating Manual
Duration (Charge
Delay)
MOP03-EN-03