Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Magstim Rapid
The customer or the user of the equipment should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-4
Electrical fast
Transient/ burst
IEC 61000-4-11
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
Power Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
Immunity
EN61000-4-3
Conducted RF
Immunity
EN61000-4-6
NOTE
U
is the a.c. mains voltage prior to application of the test level
T
© The Magstim Company Limited
2
System is intended for use in the electromagnetic environment specified below.
IEC 60601
Compliance
Test Level
± 6 kV contact
Meets requirement
± 8 kV air
± 2 kV for power
Meets requirement
supply lines
(Resetting of the
MEP Pod may be
± 1 kV for input/
experienced during
output lines
single shot mode
under these
conditions.)
± 1 kV differential
Differential mode
mode
compliant.
± 2 kV common
Common mode
mode
compliant to ± 1kV
only.
<5 % U
Meets requirement
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
) for
T
5 cycles
70 % U
T
(30 % dip in U
) for
T
25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A/m 50Hz
Meets requirement
80MHz – 2.5GHz
Coil adapter P/N
2Hz 80% amplitude
3110-00 and
modulation
Cooled Coil P/N
3380-00 between
80MHz and
100MHz may
produce system
errors setting
system into safe
mode.
0.15MHz – 80MHz
Coil adapter P/N
2Hz 80% amplitude
3110-00 and
modulation
Cooled Coil P/N
3380-00 between
30MHz and 80MHz
may produce
system errors
setting system into
safe mode.
Electromagnetic environment -
Level
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Anti surge protection needs to be
incorporated into the main supply to the
equipment if surge protection is to be
guaranteed.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Magstim Rapid
requires continued operation during power
interruptions, it is recommended that the
Magstim Rapid
interruptible power supply.
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment
Equipment should only be used in the
vicinity of other equipment compliant with
EN60601-1-2.
Equipment should only be used in the
vicinity of other equipment compliant with
EN60601-1-2.
68
Magstim Rapid²
Operating Manual
guidance
2
System
2
System be powered from an
MOP03-EN-03