Problem
The product cannot be
turned on.
Operating the remote
controller does not cause
a motorized device to
operate.
The motorized focus unit
does not operate.
The motorized focus unit
does not move down
completely.
The motorized zoom
function works
unexpectedly.
The LCD of the remote
controller is invisible.
The LED of the remote
controller is off.
The light of the LED
diascopic illumination base
cannot be adjusted.
Chapter 9
Cause
The power cable is not connected.
The power of the P2-CTLA Control Box is not
turned on or the AC adapter is not connected.
The devices are not connected or the
connections are not made correctly.
The unit is placed in escape status.
The STOP switch is pressed.
Limit setting has been made.
The ALZ function is enabled.
The backlight is turned off.
The backlight control dial is set to the
minimum level.
The LED power ON/OFF switch is set to OFF.
The control right has not been matched.
Troubleshooting
Connect the power cable properly.
Check the connections between the AC adapter
and the P2-CTLA Control Box, and then move the
power switch to the "|" position to turn on the
power.
Connect the devices correctly.
Exit the escape status.
Press the STOP switch to release the forced stop
state.
Cancel the limit setting.
Disable the ALZ function.
Turn the backlight control dial.
Turn the backlight control dial.
Set the LED power ON/OFF switch to ON.
Match the control right.