 P-DSL32 LED Diascopic Illumination Stand cable connection (Connection of the power supply to
diascopic illumination)
• When not using a control box:
Connect a 12 V AC adapter to the AC adapter input terminal on the left side of the stand's base, and then connect
the power cord.
• When using a control box:
Connect the LED brightness control terminal "CTRL" on the left side of the stand's base to the connector
"LEDDIA" on the rear side of the control box. Use the CTRL cable supplied with the control box. Connect the 24 V
AC adapter to DC24V IN of the control box, and then connect the power cord.
AC adapter input terminal
INPUT
12VDC
,
3.5A
Left side of the base
Beware of overhang of cables
When these cables are connected, the connectors (cables) overhang to the left of the base. Be careful so that they do
not interfere with other devices such as a control box or they are not accidentally touched and disconnected while
attaching devices or performing observation.
Use the power cord specified in Chapter 5, "1 Performance Properties."
WARNING
Before connecting the power cord
To prevent electric shock, be sure to turn off the power switch (press it to the "O" position) at the front of the base before
connecting the power cord to the adapter.
 Fiber attachment to the P-DSF32 Fiber Diascopic
Illumination Stand
Insert a hex driver (nominal designation: 2) in the fiber
fixing screw hole at the right rear side of the base and
loosen the internal fixing screw (M4 set screw). Insert the
fiber to the rear side of the base until it reaches the limit
and tighten the fixing screw.
Firmly insert the other end of the fiber into the fiber
attachment hole of the C-FLED2 LED Light Source for
Fiber Illuminator. See the instruction manual for the light
source for details.
Chapter 3
CTRL
CTRL
ON
OFF
LED brightness control
selection switch
LED brightness control terminal
Assembling
Rear side of the P2-CTLB
Securely attaching the fiber
43