2)
Fill in the gap between the top part and the
barrier filter with adhesive.
(4)
Bond to attach the excitation filter.
1)
Push the excitation filter into the position
shown in the figure of the bottom part until it
hits the bonding surface. Observe the
manufacturer's instructions for the orientation
of the excitation filter.
2)
Fill in the four bonding positions of the bottom
part with adhesive to bond the excitation filter
to the bottom part.
(5)
Bond to attach the dichroic mirror.
1)
Place the dichroic mirror so that its evaporated
surface contacts the bonding surface and the
side touches the abutting surface.
2)
Place the dichroic mirror at the center of the
bonding surface and place adhesive at the four
bonding positions shown in the figure.
Adhesive application
• Make sure that the adhesive does not seep in
between the dichroic mirror and the bonding
surface.
• Adhesive must not leak from the bottom part.
(6)
Assembling each part of the P2-EFLC.
1)
Securely attach the cover to the bottom part by
tightening the four screws on the sides.
2)
Securely attach the top part by tightening the
two screws on the top.
Chapter 3
Assembling
Adhesive
at the four
positions
36
Bond along the
circumference.
Barrier filter bonding - 2
Excitation filter setting
Excitation filter setting
Dichroic mirror setting
Dichroic mirror bonding
Cover attached
P2-EFLC assembly
Bond four positions
DM
Adhesive
Bottom
part
Top part attached
(complete)