Cable routing
Route the cables before attaching the cover.
(In the figure, even though the relay box is already
attached for explanatory reasons, the attachment of
the relay box is explained in step [5])
Bring out the cable from the SMZ zooming body
along the notch on the focus unit cover and connect
to the connector "ZOOM." Pass the cable from the
intelligent nosepiece through the oval hole on top of
the mount, bring it out along the notch on the top of
the left side cover, and then connect to the
connector "NP."
When the cable was connected to the motorized
focus unit, route the cable as shown in the figure.
Pass the cable from the motorized epi fluorescence
attachment through the oval hole at the top of the
mount, drop it down, and then connect to the
connector "FL."
(2)
Attach the left cover of the intelligent nosepiece.
Route the cable and replace the intelligent
nosepiece left side cover removed in step [2]-(4).
Align and fit the two protrusions on the left side
cover to the two notches in the top center of the
intelligent nosepiece, insert the two M5 hexagon
socket head fixing bolts in the fixing bolt holes at the
bottom of the cover and firmly tighten them using the
hex driver (nominal designation: 4). After the
attachment, cover the bolt holes using the rubber
caps.
5
Attach a relay box. (P2-FU Focus Unit system only)
Attach the relay box to the P2-FU Focus Unit.
The arm for attaching to the P2-FU Focus Unit and the
metal fitting for attaching to the P2-FMDN Focus Mount
are supplied with the relay box. Use the arm to attach the
relay box to the focus unit.
(1)
The orientation of the joint of the relay box and the
arm is such that the signal cables come out from the
right side (as viewed from the front) with the relay
box attached to the focus unit. Align the box with the
two attachment arm screw holes and attach it by
tightening the M5 hexagon socket head bolts
supplied with the relay box using the hex driver
(nominal designation: 4).
Chapter 3
Assembling
Cable connection to the motorized focus unit
30
FL
C
US
LISTED
4N75
INSPECTION
EQUIPMENT
ZOOM
NP
SMZ25
MADE IN JAPAN
Cable connection to the relay box
Assembling the relay box

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smz18

Table of Contents