Performance Test Software Installation
1. o install this program to the hard drive: either copy all les (except \INS ALL") to the hard
drive, or load and run the \INS ALL" program. he INS ALL program will automatically
copy all required les to your speci ed destination drive or directory (HFS, SRM or LIF
format). he INS ALL program is only contained in the LIF software version.
Note
Any directory name used must be less than 80 characters long, and preferably
less than 60 characters, for best visibility.
2. Set the computer's \Mass Storage Is" (MSI) command to the directory or drive location of
the main program le (CAL871XC).
3. Load and run this program. he program will display the initial screen with the software
part number, revision date and current time/date. If the current time or date is incorrect,
select the
ET TIME
4. Press
. Remember, the
RE UME
5. Press
to enter the con guration screen.
CONFIG
his menu allows you to enter all the speci c information that pertains to your particular
con guration (such as test equipment, addresses, and serial numbers). he con guration
le's selection descriptions and their initial default values are listed below.
Con guration Menu
In this section, the default values for the parameters listed in bold type appear in parenthesis.
The analyzer's P-IB address (716).
Please note that this is the only address that has both the Interface Select Code (ISC) and the
device number associated with it. his address can be any valid address with any valid ISC.
not
he ISC does
have to be the same as all the other test equipment (which must be all on the
same ISC). Generally the full address for the DU is set to either 716 or 816. his also sets the
ISC to 7 or 8 respectively. he factory default HP-IB device address for the analyzer is 16.
The ISC for the test equipment (7).
equipment. his is generally 7 or 8 and may be the same as that for the DU .
The power meter model number ( P 437B) and address (13).
for the power meter model number and address.
Caution
he program will accept the HP 436A power meter as a valid selection, but it is
not recommended or supported and may cause the program to halt execution.
The spectrum analyzer ( P 8566B) and address (18).
selection. Any of the listed models will suce, but keep in mind that if you do not use a
frequency counter, then the spectrum analyzer must have high degree of frequency accuracy
since it will be used as the frequency counter.
The function generator model number ( P 8116A) and address (14).
for the function generator and address. he HP 8116A is the only valid selection at this time.
1-16
Performance T ests
softkey to set it.
key can also be used as softkey #1.
RETURN
5
he rst selection is the analyzer (DU ) HP-IB address.
he second selection is the ISC for all the associated test
he next two selections are
Next is the spectrum analyzer
he next selection is