Service Kit - Agilent Technologies 8711C Service Manual

Rf network analyzers and tuner analyzer
Table of Contents

Advertisement

For disassembly procedures, see \Assembly Replacement."
Warning
Some parts in the analyzer have sharp edges. Work carefully to avoid
injury.
Before replacing an assembly, inspect the board for obvious, easy-to- x defects. Examples
include:
bent pins on ICs
no solder in holes around the edges of shields
cold solder joints

Service Kit

he service kit (HP part number 08712-60012) for the HP analyzers contains several parts that
are useful for adjusting and servicing the analyzer. he following parts are included:
An extender board assembly that allows you to operate selected internal board assemblies
outside the analyzer chassis.
An HP-IB connector extender that allows you to mate an HP-IB cable when the extender
board assembly is attached to the analyzer.
wo SMB extension cables, 12 inches long (30 cm).
A modi ed 5/16-inch wrench for better access to SMA cable nuts.
An IC extractor tool for removing ICs such as the U335 bootROM.
A 6-inch (150-mm) semi-rigid coaxial extension cable.
A non-shielded display ribbon cable used to connect the display when the A2 CPU board is
operated outside the chassis.
A voltage reference source used to perform several adjustments. he source provides
6
24 mV, and +10 V supplies.
A \minimum con guration" bootROM.
T roubleshoot ng and Block D agrams
6
0.5 V,
4-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

87128713c8714c8730a

Table of Contents