1. Disconnect the power cord.
2. Disconnect any cables from the EX DE X-input, EX DE Y-input, and AUX INPU rear
panel BNC connectors.
3. Remove nuts and washers from AUX INPU BNC connector.
4. Remove the three anchoring screws on the bottom of the analyzer. See Figure 3-6.
5. Remove the front panel assembly (see \How to Remove the Front Panel Assembly").
Remove the handle nut plate on the right side only.
6. here are two semi-rigid cables connecting the A5 assembly to the A4 assembly. Completely
loosen one nut on each cable where it connects to the A4 assembly, then separate the cable
from the A4 assembly.
Caution
Be very careful when reconnecting the semi-rigid cables (items numbered 1
and 4 in able 7-1 and able 7-2 in the \Parts List" chapter). hey are not
interchangeable. Ensure that one cable is not confused with the other. It
is possible to force the wrong cable onto the wrong connector, resulting in
damage to the connectors/cables. ighten to 10 lb-ins of torque.
7. Grasp the handle tab and pull forward to release the assembly. Once released, the assembly
will easily slide out of its cavity.