Source Power (AL ) orrection, Adjustment #104
In this procedure, the analyzer creates a table of values that corrects the source output power
over di erent frequencies and power levels. he analyzer reads values from a power meter to
determine the actual source output level. An HP 437B or HP 438A power meter is required for
this test. Other power meters will not be controlled correctly by the analyzer.
Note
If your analyzer has a step attenuator installed (HP 8730A or Option 1E1
analyzers), you need an HP 8481D power sensor to perform this adjustment

procedure.

equired Equipment
Equipment
escription
Power meter
Power sensor
High sensitivity power sensor
(for HP 8730 or Opt. 1E1
analyzers only)
10 dB attenuator*
Minimum loss pad
HP-IB cable
*The HP 8713C and HP 8714C may not require an attenuator for this adjustment.
Warmup Time
Allow the analyzer to warm up for one hour.
stimated Adjustment Time
his adjustment takes approximately ve to ten minutes to run, depending on the installed
options.
Procedure
2-22
Adjustments
Recommended Model or Part Number
50
HP 437B/438
HP 8482
HP 8481D
HP 8491 Opt. 010
N/
HP 10833
75
HP 437B/438
HP 8482
HP 8481D
N/
HP 11852B
HP 10833