Agilent Technologies 8711C Service Manual page 56

Rf network analyzers and tuner analyzer
Table of Contents

Advertisement

4. o proceed with the adjustment, disconnect the frequency counter and adapter(s) from the
RF OU PU connector.
5. Remove the handles and front panel by following these steps and referring to Figure 2-7.
a. Remove the trim strip from the handles.
b. Remove the screws attaching each handle to the analyzer.
c. Pull the analyzer toward you until it extends about two inches over the edge of the table
top.
d. Grasp and pull the front panel with two hands: one on the top-middle of the panel, and
the other on the bottom-middle of the panel.
Insert artwork here.
6. Reconnect the frequency counter and adapter(s) to the RF OU PU connector.
7. o obtain a counter reading of 500 MHz
A3 fractional-N/reference assembly (accessible through a hole in shield between J1 and J3).
8. Reassemble the analyzer.
Figure 2-7. Removing the andles and Front Panel
6
2500 Hz or better, adjust R1 on the
Adjustments
2-13