Note
If you must cycle the analyzer's line power, and you did not store the
correction constants to disk before starting this adjustment, you need to
repeat all of the adjustments done up to this point, then repeat the source
power correction. If you did save the correction constants to disk, recall
the correction constants data from disk as described in \Loading Correction
Constants from Disk" in Chapter 8, then go back to step 1 to repeat this
adjustment.
2-24
Adjustments