Performance Tests and Calibration
Adjustment Considerations

Valid adjustment reference input values

Adjustments can be performed using the following reference input values:

Table 6-5 U1211A valid adjustment reference input values

Function
Range
DC voltage
Short
400 V
1000 V
AC voltage
400 V
1000 V
AC current
40 A
400 A
1000 A
Resistance
Short
4 kΩ
400 Ω
Capacitance
400 µF
4000 µF
Diode
Short
2.000 V
102
Reference input value
SHORT
300.0 V
1000 V
030.0 V (70 Hz)
300.0 V (70 Hz)
300.0 V (2 kHz)
100 V (70 Hz)
1000 V (70 Hz)
1000 V (2 kHz)
02.00 A (70 Hz)
30.00 A (70 Hz)
030.0 A (70 Hz)
300.0 A (70 Hz)
50 A (70 Hz)
300 A (70 Hz)
SHORT
3.000 kΩ
300.0 Ω
300.0 µF
3000 µF
SHORT
2.000 V
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Short Ω and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0 Ω
0.9 to 1.1 × reference input value
User's and Service Guide