4. Connect the equipment as shown in Figure 2-4.
Insert artwork here.
5. Load and run the test software by following the procedure in \Loading and Running the
est Software" in the \Performance ests" chapter of this guide.
6. Select \Fractional-N Adjustments" from the performance tests main menu.
7. For each API adjustment, the program shows spurs three times, once for each fractional-N
VCO frequency. o minimize the spur, adjust the potentiometers shown in Figure 2-5.
You may have to compromise the three di erent adjustments of the same pot to achieve
the best overall results.
Generally, the API #1 spur is the worst. When adjusting the API #2 and API #4 spurs, their
amplitude may be so low that they are not visible.
Insert artwork here.
Figure 2-4. Setup for Fractional-N Spur Adjustments
Figure 2-5. API and 100 k z Adjustments
Adjustments
2-9