Select Adjustment Menu

Adjustments generate correction constants (CCs) which compensate for the hardware
performance of the analyzer. Correction constants are stored in segmented ash EPROMs along
with the rmware. he data is copied to RAM during power-up. Running an adjustment test
modi es the data in RAM, but it does not alter the data in EPROM.
o store the new CCs into EPROM, refer to Chapter 8, \Correction Constants and Firmware."
For details on how to perform all necessary adjustments, refer to Chapter 2, \Adjustments."
5-4
Serv ce Related Menus
Table 5-1. Analyzer Self-Tests (continued)
Prompts user to press any keys and displays information on the key
pressed. Press PRESET
Tests all locations of battery-backed SR M by writing zeroes. This
erases existing data.
Not implemented.
Not implemented.
Not implemented.
Prompts user to connect a printer or terminal, and then sends
characters to the port.
Prompts user to connect a printer, and then sends characters to the
port.
Prompts user to connect a DIN keyboard. Displays feedback on
detected key presses.
Prompts user to disconnect HP-IB cable. Tests HP-IB hardware.
Prompts user to connect a printer, and then sends characters to the
port.
Prompts user to disconnect cable and then tests the hardware.
Prompts user to ground the BNC and then tests the hardware.
Prompts user to disconnect cable and then tests the hardware.
Prompts user to ground the BNC and then tests the hardware.
escription
three times to exit.